The reported ipg and csl were received at cvrx for analysis.A visual inspection of the ipg revealed no physical damage.There was a single suture in each suture hole of the ipg indicating the ipg had been sutured into tissue and had pulled out of the tissue.The csl was received in three pieces.Growth was present on the lead, and the silicone tubing was significantly yellowed.The distal section was 7-8 inches in length, does not appear to have an insulation breach, and was twisted near the molded transition.The lead appears to be cut.The middle section of the csl appeared to be cleanly cut at one end through the tubing and coil, which is most likely associated with being cut with a tool to aid the explant.This cut matched the cut on the distal portion of the lead.The other end of the middle section of the csl appeared torn and did not exhibit any signs of a cut made by a tool.The coils of the proximal section of the csl were stretched significantly and there was significant twisting of the coils.The cause of the twisting was unable to be determined, and it wasn't clear if the twisting was a result of twiddling or caused during explant.At the conclusion of the analysis, it was confirmed that the ipg sutures had pulled out and the lead was damaged.Cvrx id# (b)(6).
|