Model Number P200DNUS |
Device Problems
Failure to Cycle (1142); Increase in Pressure (1491)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3.Date of event: unknown.No information has been provided to date.D5.Other operator of device: operator of device is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the ventilator was not cycling properly and had a large amount of ventilation.There was no patient involvement and no patient harm/adverse event reported.
|
|
Manufacturer Narrative
|
H3 and h6.Evaluation codes: updated.Device investigation: one device was received for investigation.The device was visually inspected and functionally tested.The switch cover attached to the front of the product was scratched.It was confirmed that the oscillator assembly was malfunctioning, causing the ventilation volume to exceed its set value.Since the switch cover is cracked, it may be due to the impact being applied to the main unit.The oscillator assembly and the cracked switch cover were replaced.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
|
|