SMITHS MEDICAL INTERNATIONAL LTD. PNEUPAC PARAPAC VENTILATOR W/ ALARMS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number P200DNUS
Device Problems Failure to Cycle (1142); Increase in Pressure (1491)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3.Date of event: unknown.No information has been provided to date.D5.Other operator of device: operator of device is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the ventilator was not cycling properly and had a large amount of ventilation.There was no patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
H3 and h6.Evaluation codes: updated.Device investigation: one device was received for investigation.The device was visually inspected and functionally tested.The switch cover attached to the front of the product was scratched.It was confirmed that the oscillator assembly was malfunctioning, causing the ventilation volume to exceed its set value.Since the switch cover is cracked, it may be due to the impact being applied to the main unit.The oscillator assembly and the cracked switch cover were replaced.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC VENTILATOR W/ ALARMS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL LTD.
enterprise way
luton LU3 4 BU
UK  LU3 4BU
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18759819
MDR Text Key336998304
Report Number9611178-2024-00056
Device Sequence Number1
Product Code BTL
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K020899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberP200DNUS
Device Catalogue NumberP200D/NJJP
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/25/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received03/03/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-