Intuitive surgical, inc.(isi) received the endoscope to perform failure analysis.The endoscope was analyzed and found with minor cuts to the cable insulation.The damage was located near the endoscope housing.The endoscope cable integrated connector was visually inspected and found with mechanical damage.The cable-integrated connector paddleboard exhibited mechanical damages such as scratch marks, indentations, or broken or missing pieces located at the cable's proximal end.The endoscope was evaluated and found with a camera instrument adapter defect.The endoscope was placed through a friction test, and the camera instrument adapter did not rotate properly and/or noises were heard during testing and confirmed as binding.The camera instrument adapter was removed from the endoscope housing and evaluated and found with an endoscope adapter shaft bearing contributing to friction.The endoscope was also found with damage to the button pack assembly.A visual inspection confirmed hairline cracks on the enter button lamp of the button pack assembly.Furthermore, the endoscope was found with cosmetic damage.During an inspection of the endoscope cable integrated connector, the cable integrated connector housing exhibited discoloration.
|