U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-P190
Device Problems No Display/Image (1183); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The bronchovideoscope was returned as part of the asset return process.During routine inspection of the device by olympus, no image was displayed due to a defective plug unit.There was no procedural or patient involvement associated with this event.This mdr is being submitted to capture the reportable malfunction found during the device evaluation.
 
Manufacturer Narrative
The device was returned to olympus as part of the asset return process.Upon evaluation, no image was displayed due to a defective plug unit.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information added to the following fields: h6.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that no image occured due the plug unit that may have been damaged by electrical factor.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: important information ¿ please read before use ¦precautions: caution turn the video system center on only when the endoscope connector is connected to the light source.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.¦precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system ¦inspection of the endoscopic image confirm that the wli and nbi endoscopic images are normal.5.1 troubleshooting if any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18764598
MDR Text Key336059583
Report Number9610595-2024-03801
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBF-P190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2024
Was the Report Sent to FDA? No
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/23/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-