MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L331
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Under-Sensing (1661)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2024
Event Type  malfunction  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that electrogram (egm) review was requested for this pacemaker due to several stored atrial tachy response (atr) episodes that appeared to contain noise.Review of atrial tachycardia / atrial fibrillation (at/af) burden showed chronically high atrial rates for greater than a month, however review of the atr episodes showed possible atrial flutter along with the noise.Additionally, review of data near implant shows unusual right atrial (ra) signals that may have been attributed to ra lead placement.Clinic notes indicate ra sensitivity was previously adjusted to address farfield signal oversensing, which suggests that there was now some atrial undersensing.Technical services (ts) recommended bringing the patient in-clinic for programming optimization.This pacemaker system remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI EL DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18764650
MDR Text Key336214855
Report Number2124215-2024-10754
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/19/2019
Device Model NumberL331
Device Catalogue NumberL331
Device Lot Number752320
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-