AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-1TQ290
Device Problems Leak/Splash (1354); Material Separation (1562); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2024
Event Type  malfunction  
Event Description
It was reported that partial separation of the leak detection probe interface, an e315 incompatible error and light guide rod leakage were all observed on the bronchovideoscope.The issue occurred during preparation for use for a diagnostic, bronchoscopy procedure.The procedure was completed using a similar device.There were no reports of patient harm.
 
Manufacturer Narrative
To date, the device has not been returned.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation and device evaluation results.New information added to the following fields: h3, h6 and h10 corrected field: d9 and d10 (therapy date).A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.The device was returned to olympus for inspection, and the customer's complaint of "no image (error e315)" was confirmed."the venting valve was detached" and "scope connector was leaking" reported by the customer was not confirmed.Additionally, it was found that the control section had serious liquid invasion.Based on the results of the investigation, water invaded the scope connector during user handling causing the electrical components to be damaged and the error message e315 was displayed.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: important information ¿ please read before use ¦precautions: caution turn the video system center on only when the endoscope connector is connected to the light source.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.¦precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system ¦inspection of the endoscopic image confirm that the wli and nbi endoscopic images (except bf-mp290f) are normal.5.1 troubleshooting if any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18765102
MDR Text Key337012084
Report Number9610595-2024-03811
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBF-1TQ290
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
(B)(6); (B)(6) - EVIS LUCERA ELITE VIDEO SYSTEM CENTER.
-
-