Model Number CDDRA600Q
Device Problems Failure to Charge (1085); Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  Injury  
Event Description
It was reported that a charge time out alert was received by the device.Technical support was contacted and the charge time out alert was suspected to be due to the depleted battery of the device.It is unknown if the depletion was suspected to be normal or due to premature depletion.Furthermore, the longevity of the device was inaccurate, but upon investigation, was due to normal behavior after the amount of device charging that had occurred.The device was explanted and replaced to resolve the event.The patient was stable and will continue to be monitored.
 
Manufacturer Narrative
The reported event of extended charge time was confirmed by reviewing the data in the device image.However, the extended charge time was due to excessive number of high voltage (hv) charges within a short timeframe that didn¿t leave enough time for the capacitors to properly charge and discharge.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and hv output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUTRINO NXT DR
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18766697
MDR Text Key336103263
Report Number2017865-2024-33046
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDDRA600Q
Device Lot NumberP000180440
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-