• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97715
Device Problems Overheating of Device (1437); Failure to Power Up (1476); Delayed Charge Time (2586); Battery Problem (2885); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2022
Event Type  malfunction  
Manufacturer Narrative
Continuation of d10: product id 97755 lot# serial# (b)(6) implanted: explanted: product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 97755, serial/lot #:(b)(6), ubd: , udi#: (b)(4) h3.Analysis was performed on [product id 97755 lot# serial#(b)(6) analysis found that there was a recharge antenna failure, poor recharge quality error.It was also noted the recharge antenna insulation was pulled away.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient regarding an external device.The reason for call was pt says that it takes a long time to charge ins, and says sometimes it doesn't connect at all to ins.Pt says "the other day like on friday the screen turned off completely and wouldn't turn back on until this morning.Pt says the recharger gets warm when charging ins.The issue was not resolved.A replacement rtm was sent out.No symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18767040
MDR Text Key336107086
Report Number3004209178-2024-05528
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Date Manufacturer Received03/07/2022
Date Device Manufactured11/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1535-2021
Patient Sequence Number1
Treatment
"SEE H11...."
Patient Age38 YR
Patient SexFemale
-
-