Catalog Number 395242 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/30/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
|
Event Description
|
It was reported that bd stpck q-syte wht 360deg w/o nut cap ns leaked the following information was provided by the initial reporter: this is a complaint about liquid leaking.It was reported that liquid leaks even when the three-way valve is closed.
|
|
Manufacturer Narrative
|
Our quality engineer inspected the 2 photos and 1 sample submitted for evaluation.The reported issue of leakage was not confirmed upon inspection and testing of the sample.Analysis of the sample showed that there were no observable defects present.The sample underwent leakage testing, and no signs of leakage were observed.Bd could not determine a manufacturing related root cause since the reported defect was not confirmed during the evaluation of the sample.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.We regret any inconveniences this incident may have caused you and your facility.
|
|
Event Description
|
No additional information provided.
|
|