WILSON-COOK MEDICAL INC HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC; KNQ, DILATOR, ESOPHAGEAL
|
Back to Search Results |
|
Catalog Number HBD-W-15-16.5-18 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/29/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot number was not provided.Investigation conclusion: a corrective action (capa) has been initiated to reduce occurrences of catheter cracking, splitting or breaking for hbd-w devices.The product said to be involved is included in the scope of the corrective actions.In the report it does not state if negative pressure or lubrication were applied to the balloon prior to advancing down the endoscope accessory channel.The ifu states: "to facilitate passage through the endoscope, apply negative pressure to the device." "maintain balloon deflation with negative pressure and introduce into endoscope accessory channel, advancing in short increments until balloon is completely visualized endoscopically." "apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel." prior to distribution, all hercules 3 stage wire guided balloons esophageal-pyloric-colonic are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a corrective action has been initiated in an effort to reduce occurrences of this nature.This product was manufactured prior to implementation of this corrective action.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
Event Description
|
During an unknown endoscopic procedure, the physician used a cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic.It was reported that the user had a balloon dilator that leaked fluid out the side of the catheter.Per the initial reported, the defective device it looks bent and has a hole.Our attempts to collect additional information regarding patient outcome were unsuccessful.While the complainant did not specify if the patient experienced any adverse effects or required additional medical procedures due to this event, the information able to be collected does not reasonably suggest the patient was adversely impacted.
|
|
Search Alerts/Recalls
|
|
|