• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; NO MATCH Back to Search Results
Model Number CDDRA500Q
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/05/2024
Event Type  malfunction  
Event Description
It was reported that the patient presented for implant procedure.During procedure it was noted that the atrial set screw of the implantable cardioverter defibrillator (icd) was unable to be tightened.The procedure was completed using another icd.There were no patient consequences.
 
Manufacturer Narrative
The reported field event of set screw issue was confirmed.The implantable cardioverter defibrillator (icd) was returned for analysis.Septum material was observed inside the atrial set screws hex cavity.This did not allow the set screw to be tightened and secured properly in the field.The issue was consistent with having occurred during the procedure.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18773609
MDR Text Key336296703
Report Number2017865-2024-33201
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDDRA500Q
Device Lot NumberP000166653
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-