• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) NEUTRINO NXT DR; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) NEUTRINO NXT DR; No Match Back to Search Results
Model Number CDDRA600Q
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2024
Event Type  Injury  
Event Description
It was reported, that a ventricular tachycardia (vt) episode delivering anti tachy cardia therapy (atp).That was reportedly, diagnosed inappropriately.Due to chamber onset was observed, on remote transmission.The evaluation window looked 8 seconds, prior to the first non-sinus interval average, which occurred, prior to the start of the stored electrocardiogram (egm).This was the case, due to a large number of intervals for detection.Programming changes to episodal pacing were recommended, since there was a concern about no atrial markers being shown on egm.No patient symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUTRINO NXT DR
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18773759
MDR Text Key336176360
Report Number2017865-2024-33392
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model NumberCDDRA600Q
Device Lot NumberP000129451
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-