• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 101967-931
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 01/24/2024
Event Type  Injury  
Manufacturer Narrative
A review of the device's manufacturing records revealed that the sensor lot met all requirements including sterilization requirements for release.The sensor is inserted by making a small incision and placing it under skin and potential for developing skin irritation/swelling/pain/ inflammation/infection around the insertion site is a known anticipated adverse event.The user went to a scheduled sensor removal/insertion appointment where the hcp noticed an affection at the insertion site of the old sensor.The hcp decided not to remove the sensor and to wait three weeks before attempting the removal.The new sensor was inserted into the opposite arm.No antibiotics were prescribed by the hcp.No further investigation for found necessary.
 
Event Description
On (b)(6) 2024, senseonics was made aware of an adverse event where hcp noticed an infection at the insertion site during the removal procedure.The hcp decided not remove the sensor until the infection was healed and to insert the new sensor in the opposite arm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key18773792
MDR Text Key336178261
Report Number3009862700-2024-00036
Device Sequence Number1
Product Code QHJ
Combination Product (y/n)Y
Reporter Country CodeIS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/03/2024
Device Model Number101967-931
Device Catalogue NumberFG-4500-31-302
Device Lot NumberWP09511
Was Device Available for Evaluation? No
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-