BOSTON SCIENTIFIC CORPORATION AUTOGEN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G179 |
Device Problems
Premature Discharge of Battery (1057); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the battery of this implantable device was suspected to be depleting prematurely.The estimated longevity decreased not as expected, from five to less than three months, within an unspecified time period.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.The patient is on a waitlist for the replacement procedure.This device remains in service.No adverse patient effects were reported.
|
|