LIVANOVA DEUTSCHLAND GMBH HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-85 |
Device Problem
Ambient Temperature Problem (2878)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/29/2024 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that, during a procedure, the 3t heater cooler displayed an error pump 1 is blocked (e07) on patient side.There is no report of any patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.Patient information was not provided.G.5.The heater-cooler 16-02-80 is not distributed in the usa, and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in usa.Livanova initiated an investigation if any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
Customer decided to order the stirrer motor and repair the equipment themselves.A device service history review has been performed and identified that the unit was manufactured in 2014 and one previous similar event has been reported in 2023.Based on all the above facts, it cannot be ruled out that the most likely root cause of the reported event is a jammed stirrer motor likely due to the wear of the component in contribution with the action of disinfectants during disinfection procedures and environmental condition in which the component was working, as high temperatures, humidity, and condensation.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Event Description
|
See initial report.
|
|
Search Alerts/Recalls
|
|
|