• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HL 20
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  malfunction  
Event Description
The event occurred in india during a routine check.It was reported that the error message "safety-s" was displayed on a hl20 pump.No harm to any person has been reported.Complaint id: (b)(4).
 
Manufacturer Narrative
The event occurred in india during a routine check.It was reported that the error message "safety-s" was displayed on a hl20 pump.No harm to any person has been reported.A getinge field service technician will be sent onsite for investigation of the device.As soon as new information becomes available a follow up medwatch will be submitted.
 
Manufacturer Narrative
It was reported that the error message "safety-s" was displayed on a hl20 pump.The event occurred during a routine check.No harm to any person has been reported.A getinge field service technician (fst) was sent for investigation and repair on 2024-03-19.All the connectors and boards were cleaned and refitted.No parts were replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.A similar case was assessed by the getinge life cycle engineering on 2024-03-04.Because the error "safety-s" was resolved by refitting of all circuit board connections the most probable root cause was determined as communication disturbances due to transition resistance that can arise from surface corrosion.The review of the non-conformities has been performed on 2024-04-04 for the period of 2014-11-20 to 2024-02-09.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device was manufactured on 2014-11-20.Based on the results the reported failure "error message safety-s" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18777151
MDR Text Key336333534
Report Number8010762-2024-00105
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number70102.8580
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/18/2014
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-