Model Number ZCT600
Device Problem Unstable (1667)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  malfunction  
Manufacturer Narrative
Section a2, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section b3 - date of event: unknown, as information not provided.Section d6b - explant date: not applicable, lens remains implanted.Section e1 - telephone number: (b)(6).Section h3 - other (81): the intraocular lens (iol) was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review and possible product return and evaluation, if there is any further relevant information a supplemental medwatch will be filed.Section h6 - health effect - impact code 4625: code for sutures and incision enlarged.Section h6 - health effect - clinical code 4581: no code available (incision enlarged) attempts have been made to obtain the missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that during a post operative follow up it was noticed that the toric intraocular lens (iol) is not on the desired meridian within the patient's left eye.Account indicated that directions for use were not followed and daily activities were not significant affected.Through follow-up we learned that the rotation is more than 10 degrees and that the doctor rotated the lens on (b)(6) 2024 during a revision procedure.Incision was enlarged.Suture was required.The enlargement of sutures and incision was not related to a product problem or issue.The 1 day post-operative refraction was noted as sphere +2.0 cylinder -3.5 axis 40.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key18780165
MDR Text Key336241041
Report Number3012236936-2024-00375
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeEG
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZCT600
Device Catalogue NumberZCT6000140
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1