It was reported that bd alaris pump module smartsite infusion set separated the following information was received by the initial reporter with the following verbatim: rcc received a complaint via email.Email(s) attached.We have a tubing on a chemo bag just come apart.Luckily, the nurse was in the room and the patient¿s chemo was being flushed and we did not have a spill.The nurse was changing the line over to flush and the tubing came apart and there was not a spill.I have the tubing in my office in a chemo bag but still has keytruda in the line.I do have the packaging of this line with it.Just let me know what you want me to do with it.Kim is entering a safety event because it could have been drug wasted and a spill.We have had another tubing come apart while starting infusion.I am brining it down.It came apart at a different location than the other.Saline was all that was in the line at the time.It has the same number on the package as the previous tubing.We have had 3 recent issues with our primary tubing set, bd#2426-0007.The first issue was condensation in the line with what looked to possibly contain mold.The second issue was tubing disconnecting during a chemo treatment.Today we had another incident with another set where the tubing came apart (in a different place) but with saline this time.Will you please investigate asap and let us know how to proceed? thanks so much, paula chemo incident-lot#(10)23115122 saline incident-lot#(10)23115485.
|