Model Number 131F7P
Device Problem Failure to Deflate (4060)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2024
Event Type  malfunction  
Event Description
As reported, when testing this swan ganz catheter, the balloon did not deflate at all when attempting to deflate it.Replacing the device for another one, the issue was solved.There was no allegation of patient injury.The device was available for evaluation.
 
Manufacturer Narrative
The swan ganz catheter involved in this case was received by our product evaluation laboratory for a full evaluation.The balloon inflated clear and concentric and remained inflated for five minutes without leakage.The balloon deflated within three seconds without a syringe attached.Maximum deflation time from full capacity is four seconds per specifications.All through lumens were patent without any leakage or occlusion.No visible damage was observed on catheter or returned syringe.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Manufacturer Narrative
Updated section h6 (component code) and h6 (investigations conclusions).Added information to section d4 (expiration date), h4 (device manufacturer date), h6 (type of investigation).The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.Based on further engineering investigation, the root was was unable to be confirmed through product evaluation since the affected unit was returned for evaluation and no defect was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CATHETERS
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key18786865
MDR Text Key336289751
Report Number2015691-2024-01376
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K231248
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number131F7P
Device Lot Number64836401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient Weight80 KG
-
-