Model Number ENSITE-AMP-02
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Event Description
At the start of a supraventricular tachycardia procedure, the amplifier booted to a solid amber color which caused a delay.Multiple reboots were attempted but issue persisted.The amplifier was rebooted with no connections in the front of the amplifier, but issue persisted.One final reboot was attempted with no cables except the power cable, which did not resolve the issue.The procedure was completed using a non abbott system (carto).
 
Manufacturer Narrative
Additional information: d9, g3, h2, h3 one ensite x amplifier was received for evaluation.Visual inspection revealed the front ports, rear ports, and chassis were free of physical damage.For evaluation purposes the port 3 and port 4 small form-factor pluggable (sfp) were temporarily exchanged.The amplifier was powered on and never initiated, then went to a solid orange ¿not-ready¿ light emitting diode (led) status.This indicated the amplifier did not start the power-on-self-test (post).It was noted that a nios message appeared against slot 3 cardiamp board, as well as a sideplane ¿state unknown¿ message.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the root cause was attributed to the cardiamp board in slot 3 and the sideplane.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE¿ X AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18788058
MDR Text Key336327209
Report Number2184149-2024-00039
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202066, K22
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberENSITE-AMP-02
Device Lot Number8679812
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-