BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 9552
Device Problems Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2024
Event Type  malfunction  
Event Description
It was reported that stent dislodgement occurred.The target lesion was located in the tortuous right coronary artery.After pre-dilation of the lesion, a 28 x 2.50 promus premier drug eluting stent was advanced, but failed to cross the lesion and dislodged at the tip of the guide extension catheter.The device was removed, and the procedure was completed with another of the same device.There were no patient complications or injuries reported.
 
Manufacturer Narrative
E1: initial reporter address 1: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18788670
MDR Text Key336338361
Report Number2124215-2024-11438
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9552
Device Catalogue Number9552
Device Lot Number0032279024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
Patient RaceAsian
-
-