Model Number C315HIS02 |
Device Problems
Break (1069); Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/06/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that intra-operative, strong stress on the right ventricular (rv) lead was generated due to dislodgement of the blue part at the lead inlet of delivery catheter during slitting, resulting in suspected coating damage or tortuosity. it was reported that there was a placement difficulty with the rv lead.The rv lead and the delivery catheter were attempted not used and were replaced. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary:the full delivery catheter was returned in segments and analyzed.Analysis indicated the valve of the catheter was damaged.The shaft of the delivery catheter was spiral slit.The mechanical operation of the catheter slitting was partially executed.Visual analysis of the catheter indicated damage during use.The analyst noted the catheter was received with the lead serial number lff683306v.The valve of the catheter was damaged/detached from the hub and stuck in lead body.The slitting is not on the center of the hub.Spiral slitting was observed from the hub to the shaft and stopped at 15.5 cm from the hub.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|