Catalog Number 9-PFO-3525 |
Device Problem
Air/Gas in Device (4062)
|
Patient Problems
Air Embolism (1697); Cardiac Arrest (1762)
|
Event Date 02/02/2024 |
Event Type
Death
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that on 02 february 2024, a 35-25mm amplatzer talisman pfo occluder was chosen for a patent foramen ovale (pfo) closure using a 9f amplatzer talisman delivery sheath.During procedure, the patient was under deep sedation, the physician advanced the dilator and delivery sheath over the guidewire through the pfo and into the left atrium and removed the dilator and the guidewire before connecting the loader, during backflow, trans-esophageal echocardiogram (toe) revealed bubbles in the left atrium.The physician mentioned an hypothesis of pressure inversion which led to air embolism from the delivery system.The physician attached the loader to the delivery sheath (in a wet-to-wet condition) and positioned the device, patient had cardiac arrest.The physician made several maneuvers to recover the patient, and the device was not released and got recaptured.The patient passed away.The cause of death was irreversible cardiac arrest.
|
|
Manufacturer Narrative
|
An event of air in device, air embolism (air in the left atrium), cardiac arrest, and death was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause for reported patient effects air embolism (air in the left atrium) led to cardiac arrest appears to be related to air in the device.However, the reported air in device could not be conclusively determined.It is possible due to procedural conditions (possible due to the pressure inversion which led to air embolism from the delivery system.); however, this cannot be confirmed.Per case detail, the reported death is related to irreversible cardiac arrest and related to procedure.In addition, embolism is a possible outcome of the procedure per the amplatzer talisman delivery system instructions for use.
|
|
Search Alerts/Recalls
|
|