MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Display or Visual Feedback Problem (1184); Failure to Power Up (1476); Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
|
Patient Problems
Pain (1994); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id 97745 serial (b)(6) : product type programmer, patient section d information references the main component of the system.Other relevant device(s) are: product id: 97745, serial (b)(6) , (b)(4) :analysis of the 97745 programmer (ptm) (serial number (b)(6) revealed main board no power.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient (pt) regarding an external device.It was reported that they got a lost controller replacement a couple weeks ago (only lost controller, not battery), but their current controller wouldn't do anything but just click and show faint yellow and green lights on top.Pt said they leave controller plugged in, so they thought controller was charged, but this started happening when they went to use controller this morning.Pt said they were hurting because of the controller not working and was going to have to take muscle relaxers.Patient services (ps) had pt read serial and model number of controller, but pt gave 97745 model instead of 97745nt a pt would normally get for a recent lost controller replacement.Ps consulted with repair, and repair and confirmed the serial number of the replacement controller was what the pt was using from repair in (b)(6) 2022.Pt reset controller during the call, but screen did not turn on, even after plugging in to ac power supply.
|
|
Search Alerts/Recalls
|
|
|