D9/h3 - a section of the percutaneous lead returned for evaluation.Incidental findings: visual inspection of the driveline in its returned state revealed a partial circumferential slit in the outer jacket approximately 0.25¿ from the controller connector.Manufacturer's investigation conclusion: the reported pump stop and low speed events while the patient was supported with the mobile power unit (mpu) were confirmed via the submitted system controller log files.A distal end driveline repair was performed by an abbott technical services representative on 22feb2024.The approximately 19" segment of driveline replaced by technical services was returned for evaluation, with the outer jacket, bionate, and metal-braided shielding removed approximately 14.5¿ from the controller connector.¿continuity testing was performed, and all the wires were found to be electrically intact.No shorts or discontinuities were found during electrical continuity testing.Upon disassembly, visual inspection of the underlying wires did not reveal any insulation breaches.The driveline was then submerged in a saline bath for high potential testing to check for current leakage through each wire's insulation, and a cut in the insulation of the yellow wire approximately 0.25¿ from the location where the outer jacket, bionate, and metal-braided shielding had been removed for the repair.The conductors of the yellow wire were exposed; however, they appeared clean and shiny.Additionally, no physical evidence of an electrical short was observed at the location of the breach.Due to the appearance and location of the observed cut, the yellow wire breach could not be conclusively correlated to the low speed/pump stop events recorded on the submitted log files.The patient remains ongoing on (b)(6), and no further events have been reported at this time.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The heartmate ii left ventricular assist system instructions for use is currently available and discusses damage due to wear and fatigue of the driveline and includes driveline care instructions.All heartmate ii left ventricular assist device drivelines have the potential for wire/shield breakdown to occur depending upon the length of use and movement/flexing over time.Information that covers visual/audio alarms and what action(s) should be performed when they occur is also provided in this document.The heartmate ii left ventricular assist system patient handbook is currently available and contains information on caring for the driveline.Information that covers visual/audio alarms and what action(s) should be performed when they occur is also provided in this document.No further information was provided.The manufacturer is closing the file on this event.
|