SMITH & NEPHEW, INC. JOURNEY UNI TIB TR INS SZ3-4/9MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 71436134 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/06/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, during a uka surgery, a piece of one (1) journey uni tib tr ins sz3-4/9mm broke off from the underside while removing the trial insert.The piece of plastic was accounted for retrieved using forceps.The procedure had been completed, without any delay, using the same device.No injury was reported as a consequence of this issue.
|
|
Manufacturer Narrative
|
Section h10: the associated device was returned and evaluated.The visual inspection revealed the trial has gouges, burrs and small pieces fractured off the device.The device shows significant signs of wear/usage.This inspection was conducted using a magnifying glass.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|
|