• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS MEXICO .IMPRESS®.BRAIDED.CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS MEXICO .IMPRESS®.BRAIDED.CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 44038KA2/CNB
Device Problems Material Separation (1562); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2024
Event Type  Injury  
Event Description
The account alleges that during a percutaneous transluminal angioplasty [pta] procedure, the catheter tip detached within the patient.The physician had acquired retrograde arterial access and during a guidewire exchange under fluoroscopy, the catheter tip detached.The physician successfully externalized the foreign body from the patient via cut down technique liberating the vessel.No additional patient consequences to report.
 
Manufacturer Narrative
The suspect device is expected to return for evaluation.A supplemental report will be submitted once the evaluation is complete.
 
Manufacturer Narrative
The suspect medical device was not returned for evaluation.A photograph was provided by the account.The complaint is confirmed.The root cause is attributed to the manufacturing process.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and one similar complaint for this lot number was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
.IMPRESS®.BRAIDED.CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego CA 92154
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego CA 92154
Manufacturer Contact
bryson heaton bsn.rn
1600 merit pkwy
south jordan, UT 84095
8012531600
MDR Report Key18793946
MDR Text Key336393849
Report Number3011642792-2024-00014
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K053171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number44038KA2/CNB
Device Lot NumberI2723977
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
INTERVENTIONAL GUIDEWIRE
Patient Outcome(s) Hospitalization; Required Intervention;
-
-