|
Catalog Number 8713050 |
Device Problem
Insufficient Flow or Under Infusion (2182)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
This report has been identified as b.Braun melsungen ag internal report (b)(4).3.Investigation results: 3.1 history inspection: the device history files were read and analyzed.Due to the lack of a precise error description, the history could not be analyzed extensively.The last infusions were investigated.No anomalies could be detected.(history files are attached to the pc notification) 3.2 visual inspection: a visual inspection was performed.The cover caps on the screw pillars, and the production seal on the lower housing were intact and undamaged.The device is in a clean state and no visible damage are to locate.(pictures are attached to the pc notification) 3.3 functional inspection: a functional test was performed.The device passes the self-test.A space line was inserted, and the pump identified the line, and it could be selected from the menu.It was possible to put the pump in operation.3.4 pressure inspection: in checking the downstream-sensor the electronic pressure cut-off and the mechanical pressure limitation of the device were tested, according to the requirements of the technical safety check.The electronic pressure cut-off was checked: pressure stage 2: is: 0,50 bar (should be: 0,1-0,7 bar) pressure stage 9: is: 1,13 bar (should be: 0,8-1,4 bar) the mechanical pressure cut-off was checked: pmax: is: 1,84 bar (should be: 1,8-2,5 bar) pmin: is: 1,57 bar (should be: >1,5 bar) safety clamp was checked: pmin: is: 1,76 bar (should be: >1,2 bar) the device matches the required values and standards.All measured values are within our specification.3.5 flow rate inspection: a delivery accuracy measurement according to iec 60601-2-24 was arranged.Here a nominal flow rate of 100 ml/h was chosen.The assessed average deviation "a" of the second operating hour was measured and resulted in a value of -0,94%.((accuracy of set delivery rate should be: ± 5 % according to iec/en 60601-2-24) the device matches the required values and standards.All measured values are within our specification.(see lab results attached to the pc-notification) 3.6 disassembling: during the investigation no faults could be detected, to investigate the inside of the device, only the upper housing was removed.No damage or soiling could be found.3.7 test equipment: description: typ nr.: lab.-id.-nr.Sika mh3151 qf04198 3.8 for examination used disposables: description: ref.: lot: infusomat space line 8700036sp 23m24e8st5 ---------------------------------------------------------------- 4.Judgment: 4.1 the complaint could not be confirmed.Summing up all tests, the infusomat space operates within our specification.No product deviation.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|
|
Event Description
|
As reported by the user facility information by bbm sales organization in france: non comform product administration time compared to the requested flow rate (administration twice as slow).".
|
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number 400640747.1.General information: complaint: (b)(4).---------------------------------------------------------------- 2.Information to the sample: 2.1 model: infusomat space 2.2 article number: (b)(4) 2.3 serial number/batch: (b)(6) 2.4 software version: (b)(4) 2.5 hours of operation: (b)(4) 2.6 further information: n/a ---------------------------------------------------------------- 3.Investigation results: 3.1 history inspection: the device history files were read and analyzed.Due to the lack of a precise error description, the history could not be analyzed extensively.The last infusions were investigated.No anomalies could be detected.3.2 visual inspection: a visual inspection was performed.The cover caps on the screw pillars, and the production seal on the lower housing were intact and undamaged.The device is in a clean state and no visible damage are to locate.3.3 functional inspection: a functional test was performed.The device passes the self-test.A space line was inserted, and the pump identified the line, and it could be selected from the menu.It was possible to put the pump in operation.3.4 pressure inspection: in checking the downstream-sensor the electronic pressure cut-off and the mechanical pressure limitation of the device were tested, according to the requirements of the technical safety check.The electronic pressure cut-off was checked: pressure stage 2: is: 0,50 bar (should be: 0,1-0,7 bar) pressure stage 9: is: 1,13 bar (should be: 0,8-1,4 bar) the mechanical pressure cut-off was checked: pmax: is: 1,84 bar (should be: 1,8-2,5 bar) pmin: is: 1,57 bar (should be: >1,5 bar) safety clamp was checked: pmin: is: 1,76 bar (should be: >1,2 bar) the device matches the required values and standards.All measured values are within our specification.3.5 flow rate inspection: a delivery accuracy measurement according to iec 60601-2-24 was arranged.Here a nominal flow rate of 100 ml/h was chosen.The assessed average deviation "a" of the second operating hour was measured and resulted in a value of (b)(4).((accuracy of set delivery rate should be:(b)(4) according to iec/en 60601-2-24) the device matches the required values and standards.All measured values are within our specification.3.6 disassembling: during the investigation no faults could be detected, to investigate the inside of the device, only the upper housing was removed.No damage or soiling could be found.3.7 test equipment: description: typ nr.: lab.-id.-nr.Sika (b)(4) (b)(4) 3.8 for examination used disposables: description: ref.: lot: infusomat space line (b)(4) 23m24e8st5 ---------------------------------------------------------------- 4.Judgment: 4.1 the complaint could not be confirmed.Summing up all tests, the infusomat space operates within our specification.No product deviation.Addition information: n/a.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|
|
Search Alerts/Recalls
|
|
|