BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Catalog Number D139504 |
Device Problems
Insufficient Cooling (1130); Obstruction of Flow (2423); Output below Specifications (3004)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/05/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro¿ catheter there was an irrigation issue with the catheter.It was reported that the catheter was set to 50 watts and the power would only increase up to 5 watts on the ngen monitor.The generator was set to somewhere around 40 watts prior to ablation.When the physician came on ablation, ablation was allowed but the power value only rose to 5 watts.The temperature immediately rose to 50 degrees.The blue arrows that typically indicate the up/down regulation of saline flow in qmode were not present.It seemed that there was an irrigation issue with the catheter because, we switched out the catheter and issues resolved.No troubleshooting with the pump was required.
|
|
Manufacturer Narrative
|
On 27-feb-2024, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.#(b)(4).
|
|
Manufacturer Narrative
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro¿ catheter there was an irrigation issue with the catheter.It was reported that the catheter was set to 50 watts and the power would only increase up to 5 watts on the ngen monitor.The generator was set to somewhere around 40 watts prior to ablation.When the physician came on ablation, ablation was allowed but the power value only rose to 5 watts.The temperature immediately rose to 50 degrees.The blue arrows that typically indicate the up/down regulation of saline flow in qmode were not present.It seemed that there was an irrigation issue with the catheter because, we switched out the catheter and issues resolved.Device evaluation details: the product was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.Visual inspection, temperature, impedance, and patency test of the returned device were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.The temperature and impedance test was performed and the device was found working correctly.No temperature or power issues were observed.Afterward, a patency test was performed and no irrigation issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The issues reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
|