MAUDE Adverse Event Report: BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Catalog Number D139504
Device Problems Insufficient Cooling (1130); Obstruction of Flow (2423); Output below Specifications (3004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/05/2024
Event Type  malfunction  
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro¿ catheter there was an irrigation issue with the catheter.It was reported that the catheter was set to 50 watts and the power would only increase up to 5 watts on the ngen monitor.The generator was set to somewhere around 40 watts prior to ablation.When the physician came on ablation, ablation was allowed but the power value only rose to 5 watts.The temperature immediately rose to 50 degrees.The blue arrows that typically indicate the up/down regulation of saline flow in qmode were not present.It seemed that there was an irrigation issue with the catheter because, we switched out the catheter and issues resolved.No troubleshooting with the pump was required.
 
Manufacturer Narrative
On 27-feb-2024, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.#(b)(4).
 
Manufacturer Narrative
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro¿ catheter there was an irrigation issue with the catheter.It was reported that the catheter was set to 50 watts and the power would only increase up to 5 watts on the ngen monitor.The generator was set to somewhere around 40 watts prior to ablation.When the physician came on ablation, ablation was allowed but the power value only rose to 5 watts.The temperature immediately rose to 50 degrees.The blue arrows that typically indicate the up/down regulation of saline flow in qmode were not present.It seemed that there was an irrigation issue with the catheter because, we switched out the catheter and issues resolved.Device evaluation details: the product was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.Visual inspection, temperature, impedance, and patency test of the returned device were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.The temperature and impedance test was performed and the device was found working correctly.No temperature or power issues were observed.Afterward, a patency test was performed and no irrigation issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The issues reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT MICRO¿ CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18794600
MDR Text Key336468636
Report Number2029046-2024-00660
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P210027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD139504
Device Lot Number31210029L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received03/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-