|
Model Number ATLGIQ |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
Hemorrhage/Bleeding (1888)
|
Event Date 02/12/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of an autolog iq instrument, the customer reported that the instrument was not processing correctly, and dumping too much blood to the waste bag.The instrument was used to complete the procedure.There was no adverse patient effect associated with this event.Medtronic received additional information that the blood that was sent to the waste bag was discarded.
|
|
Manufacturer Narrative
|
Correction d9: device available for evaluation? and return date updated.Correction h3: device evaluated? updated correction h11: device evaluation summary: the reported not processing correctly, and dumping too much blood to the waste bag was verified during service.The issue was resolved by cleaning the optics and adjusting the 100volt and 40 volt vr1 and vr2.Post repair testing was performed per specification.Note: the instrument was serviced in the field by a medtronic field service technician.The instrument was not returned to a medtronic facility.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|