SYNTHES GMBH UNK - SCREWS: FNS LOCKING; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
|
Back to Search Results |
|
Catalog Number UNK - SCREWS: FNS LOCKING |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Bone Fracture(s) (1870)
|
Event Type
Injury
|
Event Description
|
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2023, the patient underwent an unknown surgery with the fns implants.After the surgery, a secondary fracture occurred near the distal locking screw.On (b)(6) 2024, the fns implants were replaced with tfna agmt in a revision surgery.The revision surgery was completed successfully with no surgical delay.No further information is available.This report is for one (1) unk - screws: fns locking this is report 2 of 4 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D1, d2a, d2b, d3, d4, g4 - 510k: this report is for unk - screws: fns locking: rfna/unknown lot.Part and lot numbers are unknown; udi number is unknown.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|