Catalog Number L5C4531
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
B3: event dates: 02/2024.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that two (2) homechoice cassettes leaked an unspecified solution from a tiny hole in the tubing leading to the pull ring; further described as "solution bag spraying from the hole".This occurred one time during set up and a second time when the patient was connected for peritoneal dialysis therapy.The machine did not alarm for the leaks.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h6 and h10: h10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18798603
MDR Text Key336536159
Report Number1416980-2024-00809
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberL5C4531
Device Lot NumberDR23L08123
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/05/2024
Date Device Manufactured12/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CYCLER; UNSPECIFIED PD SOLUTION
-
-