Dhr was reviewed, and the pump passed all previous tests.There are no previous complaints on this device.Pump received 2/22/2024 and tested on 2/23/2024.Pump was given the initial complaint code of "2pum1: pump - acts like it is running but is not infusing", but the reportability was changed to "yes" since the patient was reported to have gotten no relief from the pump.Pump was tested with the following infusion flow rate accuracy testing procedure from document (b)(4).The pump was measured to have a flow rate deviation of 5.2 % which out of specification.Reported issue found, device not performing to specification.
|
On 02/16/2024, infutronix received a report that a pump ' acts like it is running but is not infusing", causing loss of infusion parameters.The infusion cannot resume without causing delay in treatment.Pump received 2/22/2024 and tested on 2/23/2024.See section h10 of this mdr for the detail finding.
|