MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number A-TCSE-F
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Endocarditis (1834); Perforation of Esophagus (2399)
Event Date 01/15/2024
Event Type  Death  
Event Description
The patient underwent an atrial fibrillation redo procedure on (b)(6) 2024.The patient presented with symptoms including vomiting, headache, and loss of consciousness.The patient died on (b)(6) 2024.The physician believes the cause of death was mitral endocarditis resulting from a possible atrio esophageal fistula.The official cause of death is reported as mitral endocarditis on native valve with cerebral, renal, splenic and coronary emboli with a positive blood culture for veillonella rogosae anaerobic organisms.It is noted that on the day of the procedure, diagnostic testing was done with ultrasound imaging and scanner and there was no evidence of endocarditis.Endocarditis was also not a pre-existing condition, the patient's medical history included rhythmic cardiomyopathy due to atrial fibrillation.There were no reported performance issues with any abbott device.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis; however, the log files from the tactisys quartz system were returned.The log file analysis concluded that the tacticath catheter performed as intended.The optical fibers met specifications, the recorded temperatures indicated cooling during rf ablation, and contact force measurements were displayed throughout the duration of the log files.Baseline was noted to not be reset after insertion and reinsertion into the patient as recommended in the tacticath se contact force ablation catheter instructions for use (ifu); however, due to unknown procedural conditions we are unable to conclusively determine the cause of the reported event.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Event Description
This report includes updated device information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18799168
MDR Text Key336460443
Report Number3008452825-2024-00101
Device Sequence Number1
Product Code OAE
UDI-Device Identifier05415067027696
UDI-Public(01)05415067027696(10)9254116(17)250228
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA-TCSE-F
Device Lot Number9254116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/28/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
-
-