Model Number LTF-S190-10 |
Device Problem
Peeled/Delaminated (1454)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned, and the evaluation found no additional reportable malfunctions other than those mentioned in b5.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was observed that during the device evaluation, the videoscope adhesive around objective lens was peeled.There were no reports of patient involvement.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the event, ¿glue of the objective lens peeled off¿, was confirmed, and the device did not meet the standard specifications.A root cause could not be identified.The probable cause was determined to be due to external factors or handling.The instruction manual identifies the following related verbiage which could have prevented the phenomenon: the inspection method for the suggested event is described in the operation manual ¿chapter 3 preparation and inspection 3.3 inspection of the endoscope.Olympus will continue to monitor field performance for this device.
|
|