MEDLINE INDUSTRIES, LP; DBD-RADIOLOGY PACK Back to Search Results
Model Number DYNJ61461C
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2020
Event Type  malfunction  
Manufacturer Narrative
It was reported that a syringe component was cracked and leaking occurred when drawing up an unspecified fluid.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A sample was returned for evaluation and a crack was noted along the top of the syringe.The likely root cause was determined to mishandling of the surgical pack at some point during its life cycle.Due to the reported problem/issue, and in an abundance of caution, this medwatch is being filed.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Event Description
It was reported that a syringe component was cracked and leaking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
DBD-RADIOLOGY PACK
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP
three lakes drive
northfield IL 60093 2753
Manufacturer Contact
nigel vilches
three lakes drive
northfield, IL 60093-2753
MDR Report Key18803059
MDR Text Key337011560
Report Number1423395-2024-00075
Device Sequence Number1
Product Code OES
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDYNJ61461C
Device Lot Number18LBJ267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2020
Date Manufacturer Received02/07/2020
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-