ETHICON ENDO-SURGERY, LLC. POWERED 60 ECHELON +, 340MM SHAFT; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number PSEE60A
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  malfunction  
Event Description
It was reported that during a splenectomy procedure, stapler would not fit and fire appropriate corresponding sized staple reload.Surgeon had to have another stapler be brought in the room.Surgeon believes that the staff could have put in the wrong reload size.No patient was harmed was reported.
 
Manufacturer Narrative
(b)(4) date sent: 2/28/2024 d4: batch # 259c02 investigation summary the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one psee60a device was returned with no visual non-conformances and without a reload present.In addition, a reload pan was found lodged inside the channel.The pan was removed from the channel and the device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.It is possible that handling by the customer could dislodge the pan.Dislodgement as a result of the removal of the reload from the device is the most likely scenario.Please reference the instruction for use for more information.Device history review a manufacturing record evaluation was performed for the finished device batch number 259c02, and no non-conformances related to the reported complaint condition were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERED 60 ECHELON +, 340MM SHAFT
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
PR  
3035526892
MDR Report Key18804345
MDR Text Key336633055
Report Number3005075853-2024-01737
Device Sequence Number1
Product Code GDW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPSEE60A
Device Lot NumberA9C45P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2024
Date Manufacturer Received02/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-