Catalog Number 09038078190 |
Device Problem
Chemical Problem (2893)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The analyzer serial number is (b)(6).The calibration and qc recovery data provided was acceptable.The field service engineer (fse) replaced the measuring cell and the bead mixer, and a precision check was performed successfully.The investigation is ongoing.
|
|
Event Description
|
There was an allegation of questionable elecsys vitamin d assay results for 1 patient sample on a cobas 6000 e601 module.The initial vitamin d result was 8.97 ng/ml.The sample was repeated on a cobas e411 analyzer and the result was 15.41 ng/ml.Clarification on what date the e411 result was from was not provided.The repeat result was deemed correct.
|
|