Model Number M00516730
Device Problem Migration (4003)
Patient Problem Dysphagia/ Odynophagia (1815)
Event Date 06/26/2023
Event Type  Injury  
Manufacturer Narrative
A2: patient's exact age is unknown.However, it was reported that the patient was under the age of 18.D4; h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.E1: the stent was repositioned by dr.(b)(6) with assistance from dr.Andrew hsu who was present in both cases (stent placement and repositioning).H6: imdrf patient code e1009 captures the reportable event of dysphagia.Imdrf impact code f23 is being used to capture the additional intervention of repositioning the migrated stent.Imdrf device code a010402 captures the reportable event of stent migration.
 
Event Description
It was reported to boston scientific corporation that a wallflex esophageal fully covered stent was implanted in the esophagus to treat an 8cm malignant stricture during an upper gastrointestinal (gi) endoscopy with stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tortuous and was dilated prior to stent placement procedure.On (b)(6) 2023, post-stent placement, the patient presented with dysphagia.Upon checking, it was noted that the stent had partially migrated proximally.The stent was repositioned using alligator forceps and the procedure was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX ESOPHAGEAL
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18810913
MDR Text Key336590272
Report Number3005099803-2024-00778
Device Sequence Number1
Product Code ESW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091510
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00516730
Device Catalogue Number1673
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/06/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-