ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-XTW
Device Problems Break (1069); Difficult or Delayed Positioning (1157); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2024
Event Type  malfunction  
Event Description
It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with a grade of 4, restricted posterior leaflet, prolapsed posterior leaflet, and dilated left atrium.A mitraclip xtw (30329r3008) was inserted without issues.However, difficulties with positioning the clip due to patient anatomy occurred.While in the mitral valve, after grasping attempts, it was observed that both grippers were not functioning as intended.It was noted that after the last grasping attempt, the grippers were not raised before opening the clip to invert.When the grippers were raised, the back side of the grippers interacted with the leaflets.Troubleshooting was performed, but the issue was unable to be resolved.Therefore, the clip was removed from the patient.While outside the patient, the clip system was examined and it was observed that both gripper lines were no longer inserted into the distal shaft of the catheter, leaving the grippers in the lowered position.After a new and better transseptal puncture, another mitraclip xtw (30329r3007) was inserted without issues and implanted on the mitral valve.To further reduce the mr, another mitraclip xtw (30329r3009) was advanced into the left ventricle (lv).However, while in the lv, the clip became caught in chordae.Troubleshooting was performed, but the issue was not able to be resolved.The clip could not be freed from chordae.Therefore, the clip was implanted on only one leaflet (the anterior leaflet) and chordae, where it was stuck, laterally to the first implanted clip.It was noted that there was some interaction of the first and second implanted clip, reducing the effect of the first clip.The procedure was completed with two clips implanted, reducing the mr to a grade of 3-4.It was noted that the mitraclip xtw (30329r3007) was stable on the leaflets.There was no clinically significant delay in the procedure.
 
Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional mitraclip device referenced in b5 is being filed under a separate medwatch report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18811694
MDR Text Key336850273
Report Number2135147-2024-00943
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2024
Device Catalogue NumberCDS0702-XTW
Device Lot Number30329R3008
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Age71 YR
Patient SexMale
Patient Weight73 KG
Patient RaceWhite
-
-