|
Model Number LF1937 |
Device Problems
Device Alarm System (1012); Energy Output Problem (1431); Difficult to Open or Close (2921)
|
Patient Problem
Hemorrhage/Bleeding (1888)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Event Description
|
According to the reporter, during a procedure, the handpiece had start-up sound and the end sound rung, but the sealing was inadequa te/partial.There was possibility of insufficient squeezing of the grip.Bleeding was observed directly from the device seal.
|
|
Manufacturer Narrative
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the sealing was inadequate/partial and there was possibility of insufficient squeezing of the grip.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a gastric cancer surgery, the handpiece had start-up sound and the end sound rung, but the sealing was inadequate/partial.There was possibility of insufficient squeezing of the grip.There was a regrasp alert.Jaws of the ligasure were cleaned after every use.Procedure was completed using the reported device since there was a timing of completing the seal.Another handpiece worked fine with the same generator.Bleeding was observed directly from the ligasure seal but blood transfusion was not necessary.Patient's progress status is good.
|
|
Manufacturer Narrative
|
(rfr/fdd: alarm activation) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|