Catalog Number EL5ML
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 2/29/2024 b3: unknown, assumed first day of month that complaint was reported d4: batch # x94x1v investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with no apparent damage.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired, and the clips did not advance into the jaw.The device was noted to have the tip of the advancer bent.The instrument was disassembled, upon disassembly the advancer was confirmed to be bent and thirteen(13) clips were found inside clip track.The event reported was confirmed and it is related to improper use of the device.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.Please reference the instruction for use for more information.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, the device didn´t close clips properly.No patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18812604
MDR Text Key336866159
Report Number3005075853-2024-01783
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberX94X1V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2024
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-