Catalog Number AMC9607C |
Device Problem
Component Misassembled (4004)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/25/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the blue clamp of a light sensitive drug set was found to be in an inverted position.This occurred during patient use.The device was used for premedication for chemotherapy.The equipment was changed to complete the procedure.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
E1: initial reporter last name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Correction: f10/h6: component codes (replace g04109 with g0405203).H10: the actual device was not available; however, three (3) retained samples were provided for evaluation.Visual inspection of the samples did not identify any abnormalities that could have contributed to the reported condition; the blue slide clamps were assembled in the correct position.The reported condition was not verified on the retained samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|