• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number AMC9607C
Device Problem Component Misassembled (4004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2024
Event Type  malfunction  
Event Description
It was reported that the blue clamp of a light sensitive drug set was found to be in an inverted position.This occurred during patient use.The device was used for premedication for chemotherapy.The equipment was changed to complete the procedure.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter last name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Correction: f10/h6: component codes (replace g04109 with g0405203).H10: the actual device was not available; however, three (3) retained samples were provided for evaluation.Visual inspection of the samples did not identify any abnormalities that could have contributed to the reported condition; the blue slide clamps were assembled in the correct position.The reported condition was not verified on the retained samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18812800
MDR Text Key336629472
Report Number1416980-2024-00849
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberAMC9607C
Device Lot Number23I05V899
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/27/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age32 YR
Patient SexFemale
-
-