• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC CANNON II PLUS REPLACEMENT HUB SET; KIT, REPAIR, CATHETER, HEMODI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC CANNON II PLUS REPLACEMENT HUB SET; KIT, REPAIR, CATHETER, HEMODI Back to Search Results
Catalog Number CAR-02400
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  malfunction  
Event Description
It was reported that: "(b)(6) 2024, the luer hub was found cracked during use on the patient.The patient was reported as fine.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The instructions for use (ifu) provided with this kit warns the user, "repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure".A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event Description
It was reported that: "10 jan 2024, the luer hub was found cracked during use on the patient.The patient was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNON II PLUS REPLACEMENT HUB SET
Type of Device
KIT, REPAIR, CATHETER, HEMODI
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNACIONAL DE CHIHUAHUA S.A. DE C.V
ave. washington 3701, edificio 4
colonia complejo industrial, las americas
chihuahua 31114
MX   31114
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18814640
MDR Text Key337637159
Report Number9680794-2024-00150
Device Sequence Number1
Product Code NFK
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K020430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCAR-02400
Device Lot Number13F21L0229
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED; NOT REPORTED
-
-