• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. MONOSYN QUI UNDYED 5/0 (1) 45CM DS19 (M); SYNTHETIC ABSORBABLE MONOFILAR SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. MONOSYN QUI UNDYED 5/0 (1) 45CM DS19 (M); SYNTHETIC ABSORBABLE MONOFILAR SUTURE Back to Search Results
Model Number C0025001
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Manufacturer Narrative
Summary of investigation: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have not received any sample for analysis, only a picture showing a sample with needle detached from the thread (thread is not wound).However, without any closed sample we cannot carry out an analysis in order to take a decision.Taking into account that no other customer complaints have been received concerning this issue for this code-batch, we consider that this is an isolated unit, but the whole batch is correct.Batch manufacturing record: reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Needle attachment strength results conducted on samples before releasing the product were 0.64 kgf in average and 0.52 kgf in minimum and fulfilled the requirements of the european pharmacopoeia (ep): 0.23 kgf in average and 0.11 kgf in minimum.Conclusion root cause analysis: it has not been possible to determine the root cause as no closed samples are received, only a picture showing an open sample with the needle detached from the thread.Final conclusion: in spite of receiving a picture showing a defective sample, without closed samples a suitable analysis cannot be performed and the case is considered not confirmed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Corrective measures: actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Event Description
It was reported an issue with monosyn quick suture.The client (veterinarian) reported that the needle detached from thread.No harm has been done to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOSYN QUI UNDYED 5/0 (1) 45CM DS19 (M)
Type of Device
SYNTHETIC ABSORBABLE MONOFILAR SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key18814783
MDR Text Key337671157
Report Number3003639970-2024-00075
Device Sequence Number1
Product Code GAM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K171001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0025001
Device Catalogue NumberC0025001
Device Lot Number122491
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-