|
Model Number 26605 |
Device Problems
Break (1069); Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/05/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that shaft break occurred.The stenosed target lesion was located in the carotid artery.A 10.0-37 carotid wallstent was advanced for treatment.However, during deployment, the stent would not open smoothly after the microcatheter was withdrawn to 30%.After angiographic observation, the whole stent was returned to the guiding catheter and withdrawn.After removal, it was found that the stent delivery shaft was fractured.The procedure was completed with another of same device.There were no complications reported and the patient status was stable.
|
|
Event Description
|
It was reported that shaft break occurred.The stenosed target lesion was located in the carotid artery.A 10.0-37 carotid wallstent was advanced for treatment.However, during deployment, the stent would not open smoothly after the microcatheter was withdrawn to 30%.After angiographic observation, the whole stent was returned to the guiding catheter and withdrawn.After removal, it was found that the stent delivery shaft was fractured.The procedure was completed with another of same device.There were no complications reported and the patient status was stable.It was further reported that the target lesion was 70% stenosed and was moderately tortuous and mildly calcified carotid artery.
|
|
Manufacturer Narrative
|
Returned product consisted of a carotid wallstent monorail.The device was received with the stent partially deployed on the delivery system.The investigator was unable to deploy the stent due to a complete separation of the outer sheath and severe sheath kinking.A visual and tactile inspection identified a complete break of the outer sheath at more than one location.The sheath of the device was also found to be severely kinked at more than one location.This type of damage is consistent with excessive force being applied to the device.A visual and tactile examination identified no issues with the tip of the device.
|
|
Event Description
|
It was reported that shaft break occurred.The stenosed target lesion was located in the carotid artery.A 10.0-37 carotid wallstent was advanced for treatment.However, during deployment, the stent would not open smoothly after the microcatheter was withdrawn to 30%.After angiographic observation, the whole stent was returned to the guiding catheter and withdrawn.After removal, it was found that the stent delivery shaft was fractured.The procedure was completed with another of same device.There were no complications reported and the patient status was stable.It was further reported that the target lesion was 70% stenosed and was moderately tortuous and mildly calcified carotid artery.
|
|
Search Alerts/Recalls
|
|
|