MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR ESSENTI
|
Back to Search Results |
|
Model Number 37601 |
Device Problems
Failure to Interrogate (1332); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/29/2024 |
Event Type
malfunction
|
Event Description
|
Information was received from a manufacturer's representative (rep) regarding an implantable neurostimulator (ins).The reason for call was caller stated they were preparing for ins replacement from patient's existing activa pc to percept rc and upon attempting to transfer settings, once it shows "communication in progress", it gives a telemetry failure with code 67b670a5 with option to end session.Caller has tried three times and has had the same result each time.Caller stated that patient has group a that is active, programmed to bipolar at low amplitude with 60 pw, 130 rate.Caller stated patient also has group b that is interleaved.Troubleshooting was not required.The issue was not resolved through troubleshooting.Tech services (tss) reviewed inputting the settings manually and would send and email with additional information to collect log files.No symptoms reported.
|
|
Manufacturer Narrative
|
Continuation of d10: product id b35300 (serial: (b)(6); product type: ; implant date n/a; explant date n/a medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received, from the manufacturer¿s representative (rep).Reported, the cause of the telemetry failure code wasn¿t determined.The issue didn¿t resolve, so the rep manually entered the dbs settings in the replacement ins.The device would not be returned, as the customer discarded it.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them, because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|