• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.BRAUN MELSUNGEN AG INFUSOMAT SPACE; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN MELSUNGEN AG INFUSOMAT SPACE; PUMP, INFUSION Back to Search Results
Catalog Number 8713050
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
As reported by the user facility information by bbm sales organization in poland: the pump gave 2 ml of blood too little in a longer time than it was programmed.Blood should have been transfused in 3 hours 35 minutes in a volume of 25 ml.The pump administered 23 ml of blood in 4 hours12 minutes.".
 
Manufacturer Narrative
This report has been identified as b.Braun melsungen ag internal report (b)(4).The complaint is under evaluation.A follow-up report will be provided after the examination results are available.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOMAT SPACE
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
B.BRAUN MELSUNGEN AG
carl- braun strasse 1
melsungen, 34212
GM  34212
Manufacturer (Section G)
B.BRAUN MELSUNGEN AG
carl-braun strasse 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
861 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key18816769
MDR Text Key336796521
Report Number9610825-2024-00135
Device Sequence Number1
Product Code FRN
Combination Product (y/n)N
Reporter Country CodePL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8713050
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/31/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexUnknown
-
-