Conclusion not yet available, evaluation in process.A follow-up report will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|
One 7f safesheath ii introducer sheath was returned under (b)(4) without the dilator.There were no other accessories.Traces of blood were found on and inside the sheath.Note: only ½ of the sheath was returned for analysis.According to the event description summary, the valve did not break apart when sheath was peeled, and the physician had to cut the valve in order to remove it from the patient.Upon receipt of the product, the sheath was already peeled apart and the hemostatic valve did not break in half as intended.Only half of the sheath was returned without any sign or remnant of the hemostatic valve.Nevertheless, it is expected that half of the valve (or at least part of it) should have been attached to the portion of the sheath that was returned.This most likely occurred because the hemostatic valve wasn't slit sufficiently enough to break apart as intended.This made it very difficult to tear the valve in half and needed to be cut away by the physician.Per manufacturing procedure (introducer set, silicone seal slitting): once the cutting cycle is complete, the part will be placed on the machine tray.Remove the seal from the tray and perform a visual inspection using a 10x microscope.Finished center helix cut - verify the helix cut is complete and the seal was severed completely by checking for helix cut on the top and bottom of seal additional inspection for split seals.If the work order is for split seals prior to providing qa with the 5 samples, visually inspect and pull test the first 3 seals per section 7.4 seal pull test.If the pull tests are within specification, continue to run the first 5 samples for qa acceptance, visually inspect and submit to qa.Production may continue and seals may continue to be manufactured while pending the completion of the qa inspection.Partial cut - if the seal has a partial cut, ensure the cut did not cut through the full width of the seal (blade did not sever the membrane).Per qa procedure (adelante s2s introducer sheath in-process and final inspection) destructive testing sampling plan ansi z 1.4, special level 4, aql 0.40 reduced break and peel test.Using samples from fit check as described in 13.3.1, manually break the sheath and hub and verify that the seal splits easily without extreme elongation.Also verify that the split cap and seal remain secure and do not break free or loosen from the sheath hub.Per ifu: flush sheath with 5cc of saline immediately before peeling sheath away in order to minimize back bleeding.Withdraw sheath and valve over the lead or catheter and from the vessel, while keeping the lead in place.Sharply snap the tabs of valve housing in a plane perpendicular to the long axis of the sheath to split the valve and peel sheath apart while withdrawing from the vessel.Returned device analysis revealed hemostatic valve of the sheath did not break apart during use.The valve was not cut sufficiently enough to tear in half as intended.Manufacturing and qa personnel have been notified of this issue to raise awareness and were informed to pay close attention to this detail.Capa was opened to investigate the root cause and implement corrective action to prevent recurrence of this issue.According to the device history record, the introducer sheaths passed all in-process and qa final inspection steps before shipping to the customer including visual, dimensional and mechanical testing.In conclusion, based on the information available at this time it is confirmed that the product failed to meet requirements.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|