BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, UNIV ENDS,; SYSTEM/DEVICE, PHARMACY COMPOUNDING
|
Back to Search Results |
|
Catalog Number H93821 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that there was particulate matter inside the packaging of a sterile repeater pump tube set.This was observed before use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information: h3, h4, h6, and h10.H4: the lot was manufactured between october 28, 2022 to october 31, 2022.H10: the actual device was received for evaluation.Visual inspection observed a large dark foreign matter found under the yellow label of the primary packaging.Magnification confirmed a dark foreign matter that was possibly of a dead insect attached to the sticky yellow label surface only and not in fluid pathway.The foreign matter was on the external surface of the primary packaging that was lifted off with the yellow label when it was peeled off.No contamination was observed inside the sterile packaging.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|